Features Oxylos®
Customised placement tools | yes |
Protection class | IP65 connection |
Type of measuring instrument | Table top |
Box | Stainless steel |
Weight | 28 kg |
Width | 560 mm |
Height | 495 mm |
Depth | 473 mm |
Power supply | 100-250VAC -50/60Hz – 63W |
Supply pressure | Air 4 to 8 Bar (dry and de-oiled) filtered at 20μm |
Connection diameter | Tube fitting 4 x 6 mm |
Memory capacity | 4 to 32 GB |
Interface | 7” colour graphic touch screen |
Ports | 1 USB 2.0 and 1 Ethernet RJ45 |
Storage conditions | 0 – 70°C/0 – 90% HR |
Terms and conditions of use | 0 – 40°C/10 – 90% HR |
Oxygen O2 measurement | Technology : Electrochemical Accuracy : +/- 1% relative Response time T90 : < 1 s Resolution : 0.01% Lifetime : 18 months in air |
Measurement of carbon dioxide CO2 | Technology: NDIR Accuracy: +/- 2% full scale Response time T90: 7s Resolution: 0.1% Service life: 10 years |
Pressure measurement | yes |
Leakage measurement | Technology: Mass flow measurement Accuracy: +/- 0.8% full scale (µm); +/- 1% of reading + 0.1 ml/min Resolution: 0.1ml/min – 1μm Measuring range: 0.1-500ml/min – 5μm to 500μm Leakage measurement pressure: Adjustable from 10 to 500 mbar +/- 1mbar Filling flow rate: 4200 ml/min mini at 20 mbar Compressed air supply: 4 to 8 Bar (dry and deoiled) filtered at 20μm Leakage and burst curves display (recordable) |
Burst test / Fatigue test | Technology : Barometric Accuracy : +/- 0,5% full scale Resolution : 1mbar Measurement range : 0,2 – 1,4 bar Burst pressure : Up to 1.4 bar Inflation flow rate : Minimum 250l/h Display of leak and burst curves (recordable) or Fatigue test Technology : Barometric Accuracy : +/- 0.5% full scale Resolution : 1mbar Measurement range : 200 – 500 mbar Pressure : 500 mbar Flow rate : 4200 ml/min Display of leak and burst curves (recordable) |
Clogged circuit detection | yes |
Pump suction control | yes |
Compensation for atmospheric pressure variations | yes |
Storing results | yes |
USB connection | yes |
Touch screen | yes |
Traceability | Format, line, operator, batch number, comment at end of test |
Remote connection | Remote control possible for remote maintenance |
Network connection (Ethernet socket) | yes |
Synchronised automatic triggering | yes |
Automatic test without septum | yes |
Automatic needle cleaning | yes |
Options | New: Continuous measurement of injection gas Device integrated in line and synchronised on conveyor; Can be interfaced with ERPs 3G router + HMI Advanced |