Features Oxylos®

Customised placement tools yes
Protection class IP65 connection
Type of measuring instrument Table top
Box Stainless steel
Weight 28 kg
Width 560 mm
Height 495 mm
Depth 473 mm
Power supply 100-250VAC -50/60Hz – 63W
Supply pressure Air 4 to 8 Bar (dry and de-oiled) filtered at 20μm
Connection diameter Tube fitting 4 x 6 mm
Memory capacity 4 to 32 GB
Interface 7” colour graphic touch screen
Ports 1 USB 2.0 and 1 Ethernet RJ45
Storage conditions 0 – 70°C/0 – 90% HR
Terms and conditions of use 0 – 40°C/10 – 90% HR
Oxygen O2 measurement Technology : Electrochemical
Accuracy : +/- 1% relative
Response time T90 : < 1 s
Resolution : 0.01%
Lifetime : 18 months in air
Measurement of carbon dioxide CO2 Technology: NDIR
Accuracy: +/- 2% full scale
Response time T90: 7s
Resolution: 0.1%
Service life: 10 years
Pressure measurement yes
Leakage measurement Technology: Mass flow measurement
Accuracy: +/- 0.8% full scale (µm); +/- 1% of reading + 0.1 ml/min
Resolution: 0.1ml/min – 1μm
Measuring range: 0.1-500ml/min – 5μm to 500μm
Leakage measurement pressure: Adjustable from 10 to 500 mbar +/- 1mbar
Filling flow rate: 4200 ml/min mini at 20 mbar
Compressed air supply: 4 to 8 Bar (dry and deoiled) filtered at 20μm
Leakage and burst curves display (recordable)
Burst test / Fatigue test Technology : Barometric
Accuracy : +/- 0,5% full scale
Resolution : 1mbar
Measurement range : 0,2 – 1,4 bar
Burst pressure : Up to 1.4 bar
Inflation flow rate : Minimum 250l/h
Display of leak and burst curves (recordable)
or
Fatigue test
Technology : Barometric
Accuracy : +/- 0.5% full scale
Resolution : 1mbar
Measurement range : 200 – 500 mbar
Pressure : 500 mbar
Flow rate : 4200 ml/min
Display of leak and burst curves (recordable)
Clogged circuit detection yes
Pump suction control yes
Compensation for atmospheric pressure variations yes
Storing results yes
USB connection yes
Touch screen yes
Traceability Format, line, operator, batch number, comment at end of test
Remote connection Remote control possible for remote maintenance
Network connection (Ethernet socket) yes
Synchronised automatic triggering yes
Automatic test without septum yes
Automatic needle cleaning yes
Options New: Continuous measurement of injection gas
Device integrated in line and synchronised on conveyor; Can be interfaced with ERPs
3G router + HMI Advanced